Skip to main content

Table 2 Maximum displacement at the osteotomy site of each model and maximum stress on the implanted bone plate and screws under different loading directions

From: Comparing the stability of a novel h-shaped bone plate and commercial endosteal plating systems in hallux valgus correction

Loading Direction on the 1st metatarsal bone

implant

Maximum displacement of head (mm)

Maximum von Mises of Plate (MPa)

Maximum von Mises of screws (MPa)

Maximum Fractional Stress between bone/screw interface (MPa) / [screw no.]

At distal

At proximal

from dorsal to plantar (flexion/extension moment)

A-Plus

0.746

447.94

694.82

70.61 [1]

36.841 [2]

VTEK (0 mm offset)

0.289

359.71

725.49

15.48 [1]

27.24 [5]

VTEK (7 mm offset)

0.358

623.21

456.67

54.91 [1]

18.41 [6]

New Design

0.278

429.51

294.97

28.21 [3]

28.33 [4]

from medial to lateral (abduction/adduction moment)

A-Plus

0.553

603.43

709.66

61.95 [1]

35.76 [2]

VTEK (0 mm offset)

0.345

567.76

520.60

15.15 [1]

67.79 [5]

VTEK (7 mm offset)

0.293

328.72

349.41

14.17 [1]

35.67 [5]

New Design

0.152

254.27

195.40

24.61 [3]

31.90 [5]